1. Open the quartz automatic double pure water distiller unit, and remove the condenser. Insert one end of the sanding port into the frosting port of the quartz flask in sequence. The inner condenser is placed inside the inner quartz flask, while the outer condenser is inserted on the outside of the quartz flask. 2. Follow the schematic diagram for hose connections. Connect the overflow hose from the first quartz flask (marked A) to the collection pool. Then, connect the once-distilled water outlet of the inner condenser to the inlet nozzle of the outer quartz flask. The top outlet hose of the inner condenser (marked C) should also be connected to the pool. The condensate feed hose of the inner condenser connects to the top water inlet of the outer condenser through hole B2. The outer condenser's condensate inlet hose exits through the rear rubber hose hole of the chassis and connects to a water valve via a three-way pipe. One of the branches passes through hole B1 and connects to the inlet of the inner quartz flask. A "water stop clamp" is placed between the hole and the three-way tube. When operating, control the flow at about 200 drops per minute using the "water stop clamp." 3. After installing all the hoses for the quartz automatic double pure water distiller, connect the secondary distilled water outlet of the outer condenser to a receiving bottle or a polyethylene container. The three-way tube is introduced into the pool through two serpent-shaped condensers, while the other branch flows into the inner quartz flask. Fill the flask halfway until water begins to flow out of the overflow port. Before starting, adjust the water stop clamp to regulate the flow rate to approximately 200 drops per minute. Ensure the water flows steadily and overflows slightly into the flask to maintain proper operation. 4. Plug in the power supply for the quartz automatic double pure water distiller and turn on the main power switch. The working indicator light will illuminate, and the quartz heating element will gradually turn red, starting the process in the inner quartz flask. Once the water in the flask begins to boil, distilled water will start dripping into the second outer quartz flask. Wait for about an hour until the outer flask begins to overflow, then turn on the second power switch to initiate the second distillation cycle. Proper installation and operation of the quartz automatic double pure water distiller are essential to ensure clean and consistent results. Always check all connections before starting the system, and monitor the water flow regularly during operation. This device is ideal for laboratories requiring high-purity water for various applications. Make sure to follow safety guidelines and maintain the equipment regularly for optimal performance. Innovative hinge damper,aluminum fence hinges,door gap buffer hinge,soft closing push open hinge,furniture hardware Foshan Nanhai Ranming Lijiali Hardware Co., Ltd , https://www.lijialihardware.com